解析儀(yi)(yi)器儀(yi)(yi)表測硫(liu)儀(yi)(yi)在化驗(yan)中出現的問題
電(dian)(dian)(dian)(dian)(dian)解池(chi)(檢測(ce)全硫含量)由有機玻璃(li)制成。在上蓋(gai)中固定有一對(dui)鉑電(dian)(dian)(dian)(dian)(dian)解電(dian)(dian)(dian)(dian)(dian)極(ji)(ji)(ji)和一對(dui)鉑指示電(dian)(dian)(dian)(dian)(dian)極(ji)(ji)(ji),上蓋(gai)與殼體用橡膠(jiao)密(mi)封圈密(mi)封。電(dian)(dian)(dian)(dian)(dian)解電(dian)(dian)(dian)(dian)(dian)極(ji)(ji)(ji)面積為1X1.5平方(fang)厘米,電(dian)(dian)(dian)(dian)(dian)解陰(yin)極(ji)(ji)(ji)置于電(dian)(dian)(dian)(dian)(dian)解池(chi)的(de)中心(xin),陽極(ji)(ji)(ji)置于電(dian)(dian)(dian)(dian)(dian)解池(chi)的(de)邊緣,以使生成的(de)碘盡快擴(kuo)散。
(1)玻璃熔(rong)板(ban)的(de)(de)透氣性(xing)(xing)。其(qi)透氣性(xing)(xing)直接影(ying)響測試結(jie)果。因(yin)此(ci),對(dui)透氣性(xing)(xing)要定期檢查(cha)。如熔(rong)板(ban)有黑色沉結(jie)物,應(ying)及時(shi)清洗(xi)(xi)。取下電解池(chi)(不(bu)需打開(kai)蓋),在電解池(chi)中(zhong)注(zhu)(zhu)入水,其(qi)水以漫不(bu)到熔(rong)板(ban)為宜。將電解池(chi)傾斜(xie)放置,用(yong)滴管(guan)往熔(rong)板(ban)的(de)(de)支管(guan)中(zhong)注(zhu)(zhu)入新(xin)配制的(de)(de)清洗(xi)(xi)液(ye)(ye)。(洗(xi)(xi)液(ye)(ye)配制:5g重鉻酸鉀和10ml水加(jia)熱溶(rong)解,待冷卻后緩緩加(jia)入100ml濃(nong)硫酸。)待洗(xi)(xi)液(ye)(ye)流凈后,重復2~3次即可除掉黑色沉結(jie)物。然后從加(jia)液(ye)(ye)漏斗注(zhu)(zhu)入自來水,用(yong)洗(xi)(xi)耳球反復沖洗(xi)(xi)至不(bu)留洗(xi)(xi)液(ye)(ye),此(ci)時(shi)熔(rong)板(ban)應(ying)“潔(jie)白(bai)如初(chu)”。
(2)電(dian)(dian)(dian)極(ji)的(de)清洗(xi)。使用(yong)中(zhong)常(chang)會遇到電(dian)(dian)(dian)極(ji)沾(zhan)污、電(dian)(dian)(dian)解液(ye)發黃,造(zao)成(cheng)測試結(jie)果(guo)不穩(wen)定(ding)(ding)或易導致過滴定(ding)(ding),因而要(yao)(yao)定(ding)(ding)期(qi)用(yong)丙酮或乙醇小心擦洗(xi)電(dian)(dian)(dian)極(ji)。清洗(xi)后(hou)用(yong)蒸(zheng)餾水沖洗(xi),電(dian)(dian)(dian)解池裝(zhuang)好后(hou)要(yao)(yao)進行檢漏實(shi)驗(yan)。具(ju)體操作是(shi),在電(dian)(dian)(dian)解池中(zhong)注(zhu)滿水,用(yong)乳膠管將電(dian)(dian)(dian)解池的(de)抽管與燒結(jie)玻璃熔板接管連起來;關閉加液(ye)漏斗(dou)的(de)活塞,打(da)開電(dian)(dian)(dian)解池的(de)放液(ye)管,如(ru)水面(mian)不下降則表明電(dian)(dian)(dian)解池不漏氣。
(3)電(dian)解(jie)池電(dian)極易出現的(de)故(gu)障與排除。
①電解(jie)液(ye)長時(shi)間發白、不(bu)電解(jie),造成此故(gu)障原因(yin)是電解(jie)電極斷線或(huo)電解(jie)池插頭接觸不(bu)良。
②電解(jie)(jie)(jie)液(ye)發黃(huang)到發紅色、一(yi)直電解(jie)(jie)(jie),造成(cheng)此(ci)故(gu)(gu)障(zhang)(zhang)的(de)原因是指示電極(ji)斷線或(huo)電解(jie)(jie)(jie)池插(cha)頭接(jie)觸不(bu)良。遇上述兩種情況,先排(pai)查(cha)外圍插(cha)頭是否(fou)潮濕,灑有電解(jie)(jie)(jie)液(ye)。如外圍正常,可檢查(cha)自動跟蹤單(dan)元(yuan)或(huo)電位(wei)指示及終點控(kong)制單(dan)元(yuan)板是否(fou)有故(gu)(gu)障(zhang)(zhang)。總之(zhi),對于所出現的(de)各類故(gu)(gu)障(zhang)(zhang)排(pai)查(cha),一(yi)定(ding)要認真細(xi)致(zhi),然后用(yong)煤(mei)標(biao)(biao)樣進行(xing)標(biao)(biao)定(ding),以保證測量結果的(de)準(zhun)確。
電(dian)解池對測硫儀(yi)的(de)使用有很大相關,我們(men)要(yao)是想提(ti)高(gao)測硫儀(yi)本(ben)身的(de)壽(shou)命就必須維護好電(dian)解池只有把她維護好了,的(de)才可以使用的(de)更長久(jiu),測定結果(guo)才會更準確。